Examples of our team’s services, from installation and setup to dental unit troubleshooting and instrument repairs. Niknam Teb Service is by your side.
Section 1
Installation & Setup
Unit micromotor systems – multi-system hoses / Bien-Air Switzerland
Installation & Setup
MC2 IR Micromotor Dental unit manufacturer: Iran (Zagros Mini Unit) City: Tehran
Installation & Setup
MCX Micromotor + OPTIMA Display Dental unit manufacturer: South Korea (Hallim) City: Rafsanjan
Multi-System Hose (Optic) Dental unit manufacturer: Iran City: Isfahan
Installation & Setup
MCX Micromotor + OPTIMA Display Dental unit manufacturer: Iran (Dentos) City: Tehran
Installation & Setup
MC2 IR Micromotor Dental unit manufacturer: SDI Brand City: Rudehen
Section 2
Instrument Repairs (Before & After)
Equipped with a specialized repair workshop for Bien-Air Switzerland, Mectron Italy, and various dental instruments and surgical motors.
Before Repair
Issue: Severe contamination in the implant contra-angle gearbox Cause: Build-up caused by incorrect lubrication
After Repair
Action: Severe contamination removed Result: Proper function restored and the issue resolved
Before Repair
Issue: Tear in the brass cap section Cause: Opening with non-standard tools
After Repair
Action: Brass cap replaced using standard tools Result: Correct and fault-free performance
Before Repair
Issue: Dent in the head of the optic contra-angle Cause: Impact, falling from height, or sudden contact with hard surfaces
After Repair
Action: Dent corrected using appropriate tools Result: Correct function restored and dent issue improved Note: In this type of damage, the repair focus is on resolving the internal problem, and visual improvement may be limited.
Before Repair
Issue: Heavy contamination in the implant contra-angle axle section Cause: Incorrect lubrication and not using high-quality lubrication sprays
After Repair
Action: Severe contamination removed and proper lubrication performed by an experienced technician Result: Correct function and reliable performance — use of high-quality oil is recommended
Before Repair
Issue: Severe contamination in the implant contra-angle cartridge section Cause: Failure to follow hygiene protocols and incorrect lubrication
After Repair
Action: Severe cartridge contamination removed and proper lubrication performed by an experienced technician Result: Correct function and reliable performance — proper lubrication training provided
Before Repair
Issue: Severe deposits under the angle water spray plate Cause: Deposits in the dental unit water line
After Repair
Action: Deposits under the water spray plate removed by an experienced technician Result: Proper water spray from the angle nozzle — installing a water filter for the dental unit is recommended
Section 3
Dental Unit Troubleshooting at the Clinic
Repair services for various dental units
Dental Unit Troubleshooting
Resolving solenoid valve issues inside the dental unit tablet
Dental Unit Troubleshooting
Repairing the dental unit pneumatic valve Replacing pneumatic system hoses
Dental Unit Troubleshooting
Resolving dental chair movement issues
Submit a Service RequestComplete the form and our support team will contact you.